Common Phenotypes and Solutions - RNA
Additional Samples or Ladder Peaks -- Small peaks next to larger ones, etc
Most Probable Causes
- Gel-dye mix expired -- Use prepared gel-dye mix within one day
- RNA sample not denatured properly -- Heat sample at 70C for 2 minutes
- Particles in tubes -- For reagent preparation, use tubes supplied with the kit (not autoclaved)
- Chip contaminated -- Wear powder-free gloves only
Dust particles in separation channels -- Don't touch the wells of the chip; clean the electrodes; load the chip immediately after
taking it out of its sealed bag
Probable Causes
- Dye agglomerates present in the gel-dye mix -- Use dye concentration according to the Reagent Kit Guide
- Chip preparation with cold reagents/chips -- Prepare a new chip
Least Probable Causes
- RNA ladder/sample degraded -- Always wear gloves when handling chips/RNA samples to prevent them from getting contaminated
Degraded RNA Ladder and/or Samples -- No or too few peaks
Most Probable Causes
- RNase contamination of the pin-set -- Decontaminate the pin-set
- RNase contamination of chips and/or reagents -- Use a new chip and/or fresh reagents; wear powder-free gloves when preparing chip
Spikes -- Sharp up and down peaks together
Most Probable Causes
- Vibration of the bioanalyzer -- Don't touch the bioanalyzer during a run; remove vibration devices from bench
- Particles in tubes -- For reagent preparation, use tubes supplied with the kit (not autoclaved)
Chip/gel-dye mix contaminated -- Prepare new chip with new gel-dye mix; wear powder-free gloves only; don't touch the underside of
the chip; don't touch the wells of the chip; clean the electrodes; load the chip immediately after taking it out of its sealed bag
Probable Causes
- Dye agglomerates present in the gel-dye mix -- Use dye concentration according to the Reagent Guide
Least Probable Causes
- Power outlett -- Install power filter
Low Sensitivity -- Entire electropherogram is small, squiggly up and down "peaks"
Most Probable Causes
- Gel-dye mix expired -- Use prepared gel-dye mix within one day
- Dye concentration too low -- Use dye concentration according to the Reagent Kit Guide
- Pipetting error during preparation of mixtures -- Check dilution procedure and calibration of pipette
- Chip pipetting error -- Pipette new chip; always insert the pipette tip to the bottom of the well when dispensing the liquid
Probable Causes
- Fingerprint on focusing lens or on the backside of the chip -- Clean lens
- Insufficient vortexing of chip -- Vortex chip for 1 minute; only use IKA shaker for chip vortexing; adjust speed to 2400 rpm
Least Probable Causes
- Autofocus or laser failure -- Check autofocus and laser using the "Hardware Diagnostics"
Baseline Noise -- Baseline very squiggly with many more acute up and down peaks
Most Probable Causes
- Fingerprint on focusing lens or on the backside of the chip -- Clean lens; do not touch the underside of the chip
Chip contaminated -- Wear powder-free gloves only; don't touch the underside of the chip; don't touch the wells of the chip; clean
the electrodes; load the chip immediately after taking it out of its sealed bag
- Vibration of the bioanalyzer -- Don't touch the bioanalyzer during a run; remove vibration devices from the bench
Probable Causes
- Dye concentration too low -- Use dye concentration according to the RNA Reagent Kit Guide
Broad Peaks -- Large "hill" or "hump"
Most Probable Causes
- Sample contaminated with genomic DNA -- Check RNA-isolation protocol; remove gDNA and perform DNase treatment
- Leaks current due to contaminated pin-set -- Clean the pin-set of the electrode cartridge
- Electrodes contaminated with RNases -- Clean electrodes with RNaseZap
Probable Causes
Clogged gasket and plastic adapter of priming station -- Check the priming station as described in the seal test of the manual;
clean/replace gasket and plastic adapter if necessary
- Wrong settings of the chip priming station -- Check if chip and base plate of priming station are in the right position
Least Probable Causes
- Dye concentration too high -- Use dye concentration according to the Reagent Kit Guide
Missing Peaks -- Noisy and non-straight baseline
Most Probable Causes
Too high salt concentration in sample -- Refer to the maximum sample buffer salt concentration as specified in the reagent kit
guide; dilute samples with deionized RNase-free water if necessary
Clogged gasket and plastic adapter of priming station -- Check chip priming station as described in the seal test of the manual;
clean/replace gasket and plastic adapter if necessary
- Wrong settings of the chip priming station -- Check if chip and base plate of priming station are in the right position
Leak current due to contaminated electrodes or wet chip surface (detergents in RNA elution buffer) -- Clean electrodes with
analysis-grade water and a toothbrush; prepare a new chip, lower vortex speed or mix samples manually
Probable Causes
- Laser broken -- Perform Laser/LED/Optics test and Autofocus test
Least Probable Causes
Autofocus failure or high voltage power supply defective -- Check autofocus and high voltage power supply by means of the
"Hardware Diagnostics"
- Sample wells filled with water -- Refer to the Reagent Kit Guide for proper preparation of the chip
Missing RNA Fragment -- Only one distinct peak and marker peak observed
Most Probable Causes
Too high salt concentration in sample -- Refer to the maximum sample buffer salt concentration as specified in the reagen kit
guide; dilute samples with deionized RNase-free water if necessary
Probable Causes
Sample degradation because of RNase-contamination of electrodes or reagents -- Clean electrodes with RNaseZap; use a new chip
and/or fresh reagents; wear powder-free gloves when preparing the chip
- Incompatible column material of RNA extraction kit -- Use RNA extraction kit of different vendor
Wavy Baseline -- Severe waviness of the baseline (much like a sinoid curve)
Most Probable Causes
Following sample contaminated with gDNA causes baseline fluctuations -- Check RNA-isolation protocol; to remove gDNA, perform
DNase treatment
Leak current due to contaminated electrodes or wet chip surface (detergents in RNA elution buffer) -- Clean electrodes with
analysis-grade water and a toothbrush; prepare a new chip and lower the vortexing speed or mix samples manually
Clogged gasket and plastic adapter of priming station -- Check chip priming station for seal test; clean/replace gasket and
plastic adapter if necessary
- Contaminated electrode cartridge -- Clean the electrode cartridge
Probable Causes
- Wrong settings of the chip priming station -- Check if clip and base plate of priming station are in the right position
Least Probable Causes
- Autofocus failure of high voltage power supply defective -- Check autofocus and high voltage power supply
Cross contamination -- Unexpected peaks
Most Probable Causes
- Contamination of pipette tips -- Use fresh tips for each pipetting step
- Contamination of electrodes -- Clean electrodes between runs
Least Probable Causes
- Leaks currents due to contaminated pin-set -- Clean the pin-set of the electrode cartridge
Late Migration of the RNA Ladder or Sample
Most Probable Causes
- Vortex speed too high -- Vortex at lower (medium) speed; for chips use only the IKA vortexer
- Leak current due to dirty electrodes -- Clean electrodes
Detergents in RNA elution buffer lowers surface tension in wells of chips leading to liquid spill on top of the chip during
vortexing -- Prepare a new chip and lower the vortex speed or mix samples manually
Probable Causes
- Vortex adapter not connected tightly -- Press vortex adapter tightly on mount; replace vortex adapter
Clogged gasket and plastic adapter of priming station -- Check chip priming station with seal test; clean/replace gasket and
plastic adapter if necessary
return to HomePage
